Sample Integrity Observation #19 Sample Require filtration receive preserved

Modified on: Thu, 20 Dec, 2018 at 8:59 AM


Purpose:  Samples that are preserved prior to filtration can no longer be analyzed for dissolved parameters due to data integrity impacts.  Preservation begins the digestion/extraction process of dissolved parameters.


Most Commonly Impacts:

  • Dissolved Metals!  Samples submitted in preserved bottles for dissolved metals cannot be analyzed for dissolved metals as the digestion process will have begun and bias data
  • If field filtration is not an option, samples can be lab filtered.  Submit the samples in an unpreserved container
  • If there is no unpreserved container in the field, rinse the preservative out of a container with DI water and extra sample before filling with sample for submission to the lab


Process:  The lab contacts you to confirm how to proceed with the sample.


Did you find it helpful? Yes No

Send feedback
Sorry we couldn't be helpful. Help us improve this article with your feedback.