Why are my detection limits raised?

Modified on: Tue, 29 Jan, 2019 at 8:13 AM


The raising of Reported Detection Limits (RDL) can occur for  a variety of reasons. 


The most common reasons are:


Concentration exceeds upper calibration limit of  the test instrument.

The most common reason to raise a detection  limit is due to an adjustment of the concentration of the sample’s test parameter so that it can fall within the calibration limit of the test  instrument. This would be necessary if the concentration of the test parameter within the received sample exceeds the upper calibration level of the  instrument. In such cases it is necessary to dilute the sample to bring it within the instrument’s calibration range. The dilution value required to bring  it within the calibration range (known as the dilution factor) is also applied  to the detection limit of the analysis.


Matrix Interference

Within the test sample there is a component  which in some fashion interferes with the ability of the analyst to reliably  quantify the parameter of interest. Often by diluting the sample, the  interference is reduced to a point where an accurate quantification can be  made. The dilution required (dilution factor) is subsequently applied to the  Reported Detection Limit (RDL).


Limited Sample 

If the amount of sample received for analysis is  less than the minimum required by the laboratory, this will be reflected in the  parameter’s Reported Detection Limit (RDL). Parameter RDLs are all based upon  on an often prescribed minimum weight or volume. 



Did you find it helpful? Yes No

Send feedback
Sorry we couldn't be helpful. Help us improve this article with your feedback.

ENVIRONMENTAL