What do the comments EMPC, NDR, DPE, EMCL on my Dioxin and Furan report mean?

Modified on: Wed, 20 Feb, 2019 at 10:42 AM


An EMPC is defined in section 11.9.5.2 of Method 8290 as the following;

 

11.9.5.2 Estimated Maximum Possible Concentration -- An estimated maximum possible concentration (EMPC) is calculated for 2,3,7,8-substituted isomers that are characterized by a response with an S/N of at least 2.5 for both the quantitation ions, and meet all of the identification criteria in Sec. 11.8.4 except the ion abundance ratio criteria or when a peak representing a PCDPE has been detected. An EMPC is a worst-case estimate of the concentration.


1) The retention time must be within -1 to +3 seconds of the corresponding isotopically labelled internal standard.

2) All ion current intensities must be > to 2.5 times the background noise.

3) The integrated ion currents for the two quantitation ions must have a ratio to each other that is between the upper and lower established limits.

 

For example;

 

Natural Abundance refers to the abundance of isotopes of a chemical element as naturally found on the planet. In our dioxin/furan analysis we use the two naturally occurring isotopes of chlorine, those being 35Cl and 37Cl. 35Cl has a mass of 34.9688 and a natural abundance of 77.8%. 37Cl has a mass of 36.9659 and a natural abundance of 24.2%.

 

If we use 2,3,7,8-TCDF as an example – the 35Cl isomer of 2,3,7,8-TCDF (C12H435Cl4O) has a mass of 303.9016 and the 37Cl isomer of 2,3,7,8-TCDF (C12H435Cl337ClO) has a mass of 305.8987.  Identification criteria #3 above translates to a theoretical ion abundance ratio of 0.77 for the two 2,3,7,8-TCDF isotopes (C12H435Cl4O & C12H435Cl337ClO). The ratio of these two ions must fall between 0.65 and 0.89 in order for the peak to be considered a “positive”. Keeping in mind that the other two criteria must also be met. 


As a result we identify the analyte as EMPC / NDR – Peak detected does not meet ratio criteria and has resulted in an elevated detection limit where NDR stands for Non Detect Ratio. This translates into the worst-case scenario of the concentration.

 

 

DPE is defined as Polychlorinated Diphenyl Ethers.


11.8.4.4 Polychlorinated diphenyl ether interferences 

In addition to the above criteria, the identification of a GC peak as a Furan can only be made if no signal having a S/N > 2.5 is detected at the same retention time (± 2 sec) in the corresponding polychlorinated

diphenyl ether (PCDPE) channel.


For example, if the Hexachlorodiphenyl ether (C12H435Cl537ClO) loses 2 chlorine atoms it will have the same molecular mass as the 2,3,7,8-Tetrachlorodibenzofuran (C12H435Cl337ClO). 

As such the de-chlorinated DPE interferes with the quantitation of the 2,3,7,8-TCD Furan and will result in a raised detection limit.


Note that both NDR or DPE are used in conjunction with EMPC to signify the reason for the Estimated Maximum Possible Concentration as either elevated due to a ratio failure or a DPE interference.

 

Accurate MassIon IDElemental CompositionAnalyte
305.8987M+2C12H435Cl337ClOTCDF
375.8364M+2C12H435Cl537ClOHxCDPE
    
339.8597M+2C12H335Cl437ClOPeCDF
409.7974M+2C12H335Cl637ClOHpCDPE
    
375.8178M+4C12H235Cl437Cl2OHxCDF
445.7555M+4C12H235Cl637Cl2OOCDPE
    
409.7788M+4C12H35Cl537Cl2OHpCDF
479.7165M+4C12H35Cl737Cl2ONCDPE
    
443.7399M+4C1235Cl637Cl2OOCDF
513.6775M+4C1235Cl837Cl2ODCDPE

 

 

EMCL is defined as Estimate Maximum Concentration Level. This is when a sample is, for the case of Dioxin and Furans, diluted beyond 20 times and the analyte is still above the concentration of the highest calibration standard. Since the analysis is done by isotope dilution no further dilution of the sample can be done as the level of the internal standard is so low that it cannot be used in the isotope dilution scheme.



Did you find it helpful? Yes No

Send feedback
Sorry we couldn't be helpful. Help us improve this article with your feedback.

ENVIRONMENTAL